
Installation
Installing the Dimension 3100 System
Rev. D Dimension 3100 Manual 57
4.3 Installing the Dimension 3100 System
4.3.1 Install the Dimension 3100 SPM Unit
1. Place the vibration isolation table, or the microscope platform, at the desired location.
2. Place the vibration isolation pad on the isolation table.
3. Place the Dimension 3100 microscope in the operating location on the vibration isolation
table or other platform.
4. Carefully remove the plastic wrap from the X-Y stage.
5. Locate the clamps (2), screws (6), and flexure (1) used to secure the chuck base to the granite
block (see Figure 4.3a).
Figure 4.3a Hardware for Chuck Securement to Stage
6. Carefully slide the chuck base off the X-Y stage.
Note: Do not stretch or bend the vacuum lines. Do not remove the vacuum lines from
the chuck base.
7. Wipe down the granite and underside of the chuck base with isopropyl alcohol. Place the
chuck base back onto the granite. Dispose of wipes in an appropriately labelled solvent-
contaminated waste container.
CAUTION: The Dimension 3100 microscope unit exceeds the two-person lift
weight limit and should be lifted with a mechanical assist. Use proper
lifting technique when removing or replacing the Dimension 3100
microscope. Improper lifting may cause muscle strain or back injury.
Clamps
Flexure
Screws
Kommentare zu diesen Handbüchern