Magnetic Force Microscopy
Basic MFM Operation
Rev. D Dimension 3100 Manual 261
Figure 15.2c Cantilever Tune for Amplitude Detection
Figure 15.2d Shift in Amplitude at Fixed Drive Frequency
4. Select an appropriate Target Amplitude (approximately 2V) using Auto Tune to adjust the
Drive Amplitude so that the RMS voltage response of the photodetector is approximately
2V before tuning. Or, exit Cantilever Tune and manually adjust the Drive Amplitude
parameter under Feedback Controls.
Note: Somewhat larger values may be beneficial if using amplitude detection.
5. Quit Cantilever Tune and return to Image Mode.
6. Under Interleave Controls set the Lift start height to 0nm, and Lift scan height to 100nm.
Note: The lift height can be optimized later.
7. Set the remaining Interleave parameters (Setpoint, Drive Amplitude, Drive frequency,
Gains) to the Main Controls values by setting the flags left of the Interleave Control
column to Off.
008
Drive Frequency
Amplitude
F
0