Head, Probe, & Sample Preparation
System Information
100 Dimension 3100 Manual Rev. D
• Load the Sample: Section 7.2.12
• Focus Surface: Section 7.2.13
• Cantilever Tune (TappingMode only): Section 7.2.14
• Set Initial Scan Parameters: Section 7.2.15
• Engage: Section 7.2.16
• Withdraw: Section 7.2.17
• Advanced AFM Operation: Section 7.3
• Stage Parameters: Section 7.3.1
7.1 System Information
7.1.1 Mouse versus Trackball
The mouse exclusively operates the NanoScope software with the exception of functions related to
direct control of the stage. These commands are only initiated with the trackball. Operator-initiated
movement of motors via the Zoom, Focus, Move SPM (Z-stage) and Move XY commands are
controlled by the trackball and its buttons. Motor operation is permitted only when the appropriate
menu enables movement.
7.1.2 Motor Interlock
SPM stage movement is software interlocked against motion if a valid sum (laser signal) is not
present. The sum signal must be between 0.5-9.5 volts, otherwise all commands which try to
initiate stage motion will not move the SPM X, Y, or Z stages. However, except for upward Z-stage
travel. This interlock reduces the possibility of inadvertently crashing the tip and scanner crystal
into the sample or stage, minimizing tip and scanner damage. If during the engage sequence the
sum signal changes to a value outside the interlocked limits, the downward motion stops,
preventing damage to scanners had the Z-stage continued downward.