System Overview
Control Station Overview
6 Dimension 3100 Manual Rev. D
1.2.4 Dimension 3100 Controller
The Dimension 3100 controller controls the vacuum and air supply and optics illumination. The
Dimension 3100 controller is controlled via a serial cable connection between it and the computer
(see Figure 1.2d). The Dimension 3100 controller features gauges on the front panel to indicate
vacuum and positive pressures. The Dimension 3100 controller channels positive pressure to the
underside of the chuck during X-Y movements, allowing the chuck to glide smoothly over the
granite.
Figure 1.2d Dimension 3100 Controller (front view)
CAUTION: Do not turn the Dimension 3100 controller on its side. Turning the
controller on its side activates an alarm and turns the vacuum pump
off.
ATTENTION: Afin d'éviter de sérieux dommages mécaniques, ne tournez pas le
boîtier de contrôle sur le côté. Tourner le boîtier sur le côté
déclencheune alarme et coupe la pompe à vide.
VORSICHT: Um ernsthafte mechanische Beschädigungen zu vermeiden stellen Sie
die Dimension Control Box bitte nicht auf die Seite. Dies würde einen
Alarm auslösen und die Vakuumpumpe ausschalten.
CAUTION: The Dimension controller features a special thermostat that sets off an
alarm if the controller overheats (> 40° C). The Dimension controller
will overheat if the controller ventilation holes are blocked or if the
controller is exposed to heat from an outside source.
NanoScope®
Dimension
TM
3100 Controller
Air Pressure
Chuck Vacuum Vacuum Pump
Manual
Light Control
Status
1 2 3 4 5 6 7 8 9
3908
Seitenansicht 25
1 2 ... 21 22 23 24 25 26 27 28 29 30 31 ... 375 376

Kommentare zu diesen Handbüchern

Keine Kommentare