Electric Techniques
Electric Force Microscopy
278 Dimension 3100 Manual Rev. D
Setting Analog 2 in the Software
If you have chosen a configuration where Analog 2 applies to the tip or sample, you must enable
Analog in the software. The Analog 2 parameter appears in both the Feedback Controls and
Interleave Controls panels.
Note: For Version 4.23 and lower, Analog 2 appears only in the Feedback Controls
panel.
Instructions for Enabling Analog 2 in Software
1. Select Di > Microscope Select > Edit > Advanced, and set Analog 2 to User defined.
2. Click OK to exit both dialog boxes.
Note: For software versions 4.23 and lower, select Microscope > Calibrate >
Detector to display the Detectors Parameters window. Switch the Allow in
attenuation field to Disallow.
Instructions for Disabling Analog 2 in Software
1. Select Di > Microscope Select > Edit > Advanced, and set Analog 2 to Atten switch.
2. Click OK to exit both dialog boxes.
Note: Remember to set Allow in attenuation to Allow when finished. For all other
configurations it should be left on Allow.