Magnetic Force Microscopy
Basic MFM Operation
Rev. D Dimension 3100 Manual 263
Phase Detection
Set the Channel 2 image Data type to Phase; Z range to 3 degrees; and Line direction to
Retrace.
Note: It is important that the Scan direction be set to Retrace for both the main and
interleave scans. If instead it is set to Trace, a band may appear along the left
side of the images due to the time taken for the tip to move between the surface
and the lift scan height.
Amplitude Detection
1. Set the Channel 2 image Data type to Amplitude, Z range to 1nm, and Line direction to
Retrace.
2. Change Interleave mode to Enable to invoke LiftMode.
3. Set the Channel 2 Scan line to Interleave to display the interleaved data.
Note: This can only be done after Interleave mode is Enabled. A magnetic force
gradient image similar to that shown on the right of Figure 15.2e should appear
as the Channel 2 image. The alternating dark and light stripes represent the
recorded magnetic information, signifying a varying resonant frequency and
magnetic force gradient on the tip.
Note: Keep the Setpoint as large as possible while consistent with a good image.
Wider scans (> 25µm) will reveal separate tracks in which the magnetic stripes
are at different angles
Frequency Modulation
With the Extender Electronics Module, it may be desirable to use frequency modulation. This
activates a feedback loop which modulates the Drive Frequency to keep the cantilever phase lag at
90 degrees relative to the drive, corresponding to resonance. The frequency Data Type displays the
resulting shift in Drive Frequency in Hz, and gives the most direct, quantitative image of force
gradients.
To enable frequency modulation, follow the procedure above for obtaining an MFM image with
phase detection with the following modifications:
1. In the Interleave Controls panel, set the Input Feedback to Frequency. Turn on this
parameter by selecting the gray button next to Frequency. When selected, the gray button
becomes green, indicating the control is on.
2. Switch the Channel 2 image Data type to Frequency. Try a Data scale (frequency shift) of
approximately 10 Hz.
3. Select Other Controls, then adjust the frequency modulation gains: