Calibration
Calibrating Z
332 Dimension 3100 Manual Rev. D
6. If the Z Center Position value is not close to zero, use the Realtime > Motor > Tip Up and
Tip Down buttons to adjust.
17.8.2 Capture and Correct an Image
1. Capture an image by selecting Capture in the Realtime menu, or click on the CAPTURE
icon. When the image is captured, go to the Offline screen.
2. Remove all tilt and scan line errata from the image by selecting Offline > Modify > Flatten
Manual.
3. Set the Flatten order parameter in the dialog box to 1.
4. Go to the display screen and draw a stopband over the pit as shown in Figure 17.8b.
Figure 17.8b Draw a Stopband
5. Click Execute to complete the flattening procedure.
6. Quit the dialog box.