Electric Techniques
Electric Force Microscopy
286 Dimension 3100 Manual Rev. D
Analog 2 Voltage Applied to the Tip (With Extender Electronics)
Notice that the jumper configuration in Figure 16.2k connects the Analog 2 signal from the
NanoScope III controller (± 12 V range) to the tip, and is exactly the same as the jumper
configuration shown in Figure 16.2j, the standard configuration as shipped from the factory.
Remember to enable the Analog 2 voltage line as described in Section 16.2.2.
Figure 16.2k Jumper Configuration for Application of Voltage to Tip
Chuck Ground/Bias
Unused
Auxiliary 2 Output
Ground
STM Signal
Indicates Jumpers
Unused
Ground
Ground
AFM Tip
Sample Chuck
Analog 2 or
Gnd/OSC + DC signal
Sample
Tip
Analog 2
Kommentare zu diesen Handbüchern